Medilap 400 Primo

MEDILAP 400 PRIMO+

Medilap 400 Primo electro surgical unit consist tissue impedance monitoring system, self adjustment of power output to reduce the continuous power setting for different tissue needs, to provide user consistent performance with all digital display of power and modes.

INPUT POWER SUPPLY:
Input power180‐270V AC
Current Fuse7 Amps.
Power consumption750 VA Max.
OUTPUT CONTROL:All digital and push switch control in all modes.
SAFETY FEATURES:
Unit ConstructionAccording to IEC‐60601‐2‐2 & 60601‐1‐1.
Protection ClassII‐B.
Unit TypeCF.
OutputAll Floating Output.
Nominal Frequency350 KHz
High Frequency Leakage Current<150 μA r.m.s in all output modes.
Low frequency Leakage Current<10 μA Max
AUDIO LEVEL:All isolated tone indications to differentiate the mode of application and adjustable in 4 stages.
WEIGHT6.5Kg.
DIMENSION390X130X340 (LXHXW) mm.

 

 

  • A tissue sensing technology used for all applications to deliver user consistent performance.
  • Reduce the frequent adjustment of power setting as the tissue changes.
  • Reduce the risk of collateral tissue damage due to sparkles and charring free power.
  • Minimize the patient recovery time by performing precise and clean cutting and coagulation.
  • Minimum RF distortion in other O.T. Equipment.
  • Patient plate fault monitoring system.
  • Auto monitoring facility, displays different error codes if any error occurs during use.
  • Randomized form of spray mode.
  • Two independent power activation of monopolar output, hand switch operation and foot switch operation.
  • Separate bipolar generator with foot switch operation or hand (auto) operation, eliminates the need of foot switch while using auto mode.

 

MODESub ModeMax.Rated
Load Ω
Max. RF
Power
(watts)
Mono polar
Cut
Pure
Blend
Endo
500
500
500
400 W
350 W
250 W
Mono polar
Coag
Soft
Force
Spray
500
500
500
150 W
200 W
125 W
BipolarMicro
Macro
Bicut
100
100
100
80 W
100 W
150 W